Specification
Manufacturer: BÄHR
Technical specifications:
Quenching and deformation dilatometer BÄHR DIL 805 A/D |
---|
Technical data: - Temperature range: -130 - 1180 °C
- maximum heating rate: 400 K/s
- maximum cooling rate: 250 K/s (hollow samples), 50 K/s (solid samples)
- Cooling media: N2, Ar, He
- Forming dilatometry
- no forming below 900 °C
- max. heating and cooling: 50 K/s
- Forming force: 15 kN
- Sample geometry for quenching to room temperature
- Hollow samples: 7×4×3 mm³ (L×Øoutside×Øinside)
- Full cylinder: 7×4 mm² (L×Øoutside) up to 50 K/s cooling rate
- Sample geometry for quenching down to -130 °C
- Hollow samples: 10×4×2 mm³ (L×Øoutside×Øinside)
- Sample geometry for forming above 900 °C
- Full cylinder 10×5 mm² (L×Øoutside)
|
Typical applications: - Determination of phase transformation temperatures
- Preparation of continuous cooling transformation (CCT) and time temperature transformation (TTT) diagrams and time temperature austenitisation (TTA) diagrams
- Determination of the Curie temperature
|
Contact:
Dr.-Ing. Marco Wendler
Leipziger Straße 34 (Ledebur-Bau), Room 207
+49 3731 39-3605
Marco [dot] Wendler [at] iest [dot] tu-freiberg [dot] de (E-Mail)